The ASHRAE 110 standard calls for 4.0 LPM release rate. The release rate is a combination of the upstream pressure and the porosity of the Bronze Sintered and the size of the critical orifice. But what should the upstream pressure be? We will add a digital PSI meter to your ejector and digitally calibrate the flow to exactly 4.0LPM.